Rise Time Accuracy - Tektronix 2247A Operator's Manual

Portable oscilloscope
Hide thumbs Also See for 2247A:
Table of Contents

Advertisement

Performance
Check
Procedure
Disconnect
the test equipment
from the 2247A.
Rise
Time
Accuracy
Set:
VERTICAL
MODE
CH
1 (CH 2
off)
CH 1 VOLTS/DIV
0.2Vv
A SEC/DIV
0.2 ps
A SLOPE
St
Press the COUNTER/TIMER
button.
Select FREQ.
Adjust
the sine-wave
generator
for five divisions
of signal.
NOTE
A 1 MHz sinewave has a 10-90% rise time of 295.167 ns; 20-80%
rise time is 204.833 ns.