4.4 Judgment Setting

4.4 Judgment Setting

Overview
This is a function setting related to judgment results of the measured value.

4.4.1 HIGH Set Value

Overview
This function allows you to optionally set the upper limit value (HIGH set value) used to judge
the measured value.
If the measured value exceeds the HIGH set value, the judgment is output as HIGH.
Setting method
● Be sure to set a HIGH set value that is greater than the LOW set value. If the value is not
greater than the LOW set value, an error will occur.
● Set this so that the difference between the HIGH and LOW set values is more than twice
the hysteresis. If set to not more than twice the hysteresis, an error will occur.
● You can set HIGH and LOW set values by teaching.
For details, refer to
● Hold down the <ENTER> key for 2 seconds on the setting screen, and the setting change is
suspended and the screen returns to the previous display.
● In internal processing, comparisons are performed in units of 0.01 µm or less. (Common to all
distance types)
4-26
"4.4.4 Teaching
Mode".