Levelflex M FMP41C 4...20 mA HART
3.2
Wiring in T12 housing
1.
Unscrew housing cover (1).
2.
Insert cable (2) through gland (3).
3.
Connect cable screen to the grounding
terminal (4) within the terminal
compartment.
4.
Make connection (see terminal
assignment).
5.
Tighten cable gland (6).
Max.torque 10...12 Nm!
6.
Screw on housing cover (1).
7.
Switch on power supply.
3.3
Terminal assignment
Endress+Hauser
2
2-wire-version
power
test sockets
(output current)
1 2
3 4
L- L+
I+
I-
3
4
1 2 3 4
L00-FMP40xxx-04-00-00-xx-002
plant ground
L00-FMP4xxxx-04-00-00-en-009
Wiring
1
11
Need help?