Wiring In T12 Housing; Terminal Assignment - Endress+Hauser Levelflex M FMP40 Interface Brief Operating Instructions

Hide thumbs Also See for Levelflex M FMP40 Interface:
Table of Contents

Advertisement

Wiring
3.2

Wiring in T12 housing

1.
Unscrew housing cover (1).
2.
Insert cable (2) through gland (3).
3.
Connect cable screen to the grounding
terminal (4) within the terminal
compartment.
4.
Make connection (see terminal
assignment).
5.
Tighten cable gland (6).
Max.torque 10...12 Nm!
6.
Screw on housing cover (1).
7.
Switch on power supply.
3.3

Terminal assignment

12
2-wire-version
power
communication
resistor
Ω
(> 250
)
4...20 mA
1 2
L- L+
Levelflex M FMP40 4...20 mA HART
3
2
4
alternatively
Commubox
FXA191/195
DXR375
test sockets for
testing of the signal
current
plant ground
3 4
I+
I-
1 2 3 4
L00-FMP40xxx-04-00-00-xx-002
L00-FMP4xxxx-04-00-00-en-005
Endress+Hauser
1