English
Service
Dräger recommends that all service activities are performed by
DrägerService and that only authentic Dräger repair parts are
used.
WARNING
Risk of faulty components
Device failure is possible due to wear or material fatigue of
the components.
To maintain the proper operation of all components, this de-
vice must undergo inspection and preventive maintenance
at specified intervals.
Checks
Function and condition test for:
–
Double castors with locking brake
–
Swivel arm for patient monitor
–
Device holder/carrying system holder
–
Bed coupling
–
Hook-and-loop straps on gas cylinder holder
Technical data
Ambient conditions during transport and storage
Temperature
Atmospheric pressure
Relative humidity
Ambient conditions during operation
Temperature
Atmospheric pressure
Relative humidity
Dimensions (W x H x D)
Dimensions with bar (W x H x D)
Weight
For power socket strips in combination with the power socket strip
attachment kit
IP classification
20
Interval
Every 2 years
–20 °C to +60 °C (–4 °F to +140 °F)
500 to 1060 hPa (7.3 to 15.4 psi)
5 to 95 %, no condensation
+5 °C to +40 °C (+41 °F to +104 °F)
500 to 1060 hPa (7.3 to 15.4 psi)
5 to 95 %, no condensation
577 mm x 1400 mm x 677 mm (22.7 in x 55.1 in x 26.7 in)
577 mm x 1850 mm x 677 mm (22.7 in x 72.8 in x 26.7 in)
approx. 59 kg (130 lb)
IP21
Personnel responsible
Service personnel
Instructions for use Oxylog Trolley