Technical Data - Dräger Aquapor H300 Instructions For Use Manual

Table of Contents

Advertisement

12. Technical data

Compliance of the mains voltage with the operating voltage indicated on the rating plate must be checked before
operation�
Dimensions
Weight
Classification
Electrical data
Operating data
Height:
Width:
Depth:
Aquapor H300 humidifier:
Device (protection class according to IEC 60601) class II
Type BF applied parts:
- Heated breathing circuit
- Temperature probe
Degree of protection by housing IP22 (protected against solids with a diam-
eter of 12�5 mm or more, protected against insertion of fingers, protection
against vertically dripping water if the housing is at an incline of up to 15°)
Operating voltage:
Mains frequency:
Power consumption:
Heating plate:
Inspiratory and expiratory
hose heater:
Warm-up time:
Recommended flow rate:
Humidifier system output:
Maximum operating pressure:
Continuous noise:
Sound pressure level of the alarms:
Max� water volume:
170 mm
145 mm
200 mm
Approx� 2�3 kg
Approx� 2�5 kg
incl� supplied accessories
Aquapor H300
220 V to 240 V
Aquapor H300
110 V~ to 120 V~
50 Hz/60 Hz
max� 315 VA
170 W
22 V, 30 W
max� 30 min�
2 to 80 L/min
>33 mg/L
in the range of 2-60 L/min
>12 mg/L
in the range of 2-80 L/min
200 mbar
1
<50 dB (1 m)
max� 65 dB
200 mL
41