Ametek JOFRA DLC-155 User Manual page 17

Dynamic load compensation
Table of Contents

Advertisement

The following procedure is repeated for each
temperature:
1. Set the SET temperature.
2. Place the reference temperature probe in the
bottom of the insert.
Wait until the temperature have stabilized and the
DLC-reading is 0.00°C. If the DLC-reading
continuously shows a positive value at -30°C
(RTC-156) / -45°C (RTC-157) / -22°C (RTC-158)
/ -100°C (RTC-159) / -30°C (RTC-168) / -45°C
(RTC-187), increase the SET-temperature in
steps of 1°C until the DLC-reading displays
0.00°C.
3. Record the temperature of the reference probe in
the excel spreadsheet.
4. Move the reference probe to the upper position
60 mm (RTC-156/157/158/159/168/187/250) or
80 mm (RTC-700).
5. Wait until the temperature have stabilized and the
DLC-reading is 0.00°C.
6. Record the temperature of the reference probe
and the DLC-probe in the excel spreadsheet.
7. Go to next SET-temperature.
After the last STEP is completed:
8. Check that all values in the calculated curve fit
are less than the values specified in the table. If
this is OK, the new coefficients can be
downloaded using the CON050 software.
127414
06
2021-10-29
Page 16