Ametek JOFRA DLC-155 User Manual page 22

Dynamic load compensation
Table of Contents

Advertisement

Certificate:
If the DLC probes are supplied with a certificate, the
calibration is carried out as recommended below
according to the ITS 90 temperature scale.
The DLC-155/158 probe is as standard calibrated in
the range :
–45°C to 155°C (-49°F to 311°F).
The DLC-159 probe is as standard calibrated in the
range :
–100°C to 155°C (-148°F to 311°F).
The DLC-168 probe is as standard calibrated in the
range :
–45°C to 165°C (-49°F to 329°F).
The DLC-187 probe is as standard calibrated in the
range :
–45°C to 180°C (-49°F to 348°F).
The DLC-250 probe is as standard calibrated in the
range :
28°C to 250°C (82°F to 482°F).
The DLC-700 probe is as standard calibrated in the
range :
33°C to 700°C (91.4°F to 1292°F).
127414
06
2021-10-29
Page 21