12.2 Information on electromagnetic compatibility (EMC)

EIZO monitors were designed for the display of images and normal monitor operation.
WARNING
Special EMC provisions are required for use of the CuratOR LX1910. Installation, assem-
bly, and use must take place in compliance with the following instructions.
• Only use the cables included in the scope of delivery or recommended by the manufac-
turer. The use of other cables can result in increased electromagnetic radiation and re-
duced electromagnetic interference immunity of the device, as well as improper use.
Cable length: max. 3 m
• The monitor should not be placed on other electrical devices or positioned in their im-
mediate vicinity. If electrical devices have to be operated on or in the immediate vicinity
of one another, the monitor or system must be monitored to ensure proper operation for
the defined configuration.
• When using a portable RF communications device, maintain a distance of at least 30
cm from all parts of the monitor, including cables. Otherwise, problem-free function of
the device cannot be guaranteed.
• Persons connecting additional devices to the signal input or output when configuring a
medical system are responsible for ensuring compliance with standard IEC/EN
60601-1-2.
Electromagnetic radiation
The CuratOR LX1910 is intended for use in the electromagnetic environments noted below.
Customers and users of the CuratOR LX1910 have to ensure that the device is used in such an envi-
ronment.
Radiation test
RF radiation
CISPR11/EN 55011
RF radiation
CISPR11/EN 55011
GB9254
Harmonic currents
IEC/EN 61000-3-2
GB17625.1
Voltage fluctuations/
flicker
IEC/EN 61000-3-3
CuratOR LX1910
Instructions for Use, 08/2022
12.2 Information on electromagnetic compatibility (EMC)
Conformity
Information regarding the electromagnetic environment
Group 1
The CuratOR LX1910 uses RF radiation for internal operation
only. For this reason, the RF radiation is very low and is
therefore unlikely that the monitor will cause interference in
electronic devices in the immediate vicinity.
Class B
The CuratOR LX1910 is approved for use in a number of en-
vironments. This includes residential areas and areas con-
nected directly to the public low-voltage grid, such as private
homes.
Class D
fulfilled