EMC Information

Essential performance of RadiForce series is to display images and operate functions normally.
CAUTION
The RadiForce series requires special precautions regarding EMC and need to be installed, put into
service and used according to the following information.
Do not use any cables other than the cables that provided or specified by us.
Using other cables may cause the increase of emission or decrease of immunity.
Do not put any portable and mobile RF communications equipment close to the RadiForce series.
Doing so may affect the RadiForce series.
The RadiForce series should not be used adjacent to or stacked with other equipment. If adjacent or
stacked use is necessary, the equipment or system should be observed to verify normal operation
in the configuration in which it will be used.
Guidance and manufacturer's declaration - electromagnetic emissions
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Emission test
Compliance Electromagnetic environment - guidance
RF emissions
Group 1
EN55011
RF emissions
Class B
EN55011
Harmonic emissions
Not applicable
EN61000-3-2
Voltage fluctuations /
Complies
flicker emissions
EN61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The RadiForce series is intended for use in the electromagnetic environment specified below.
The customer or the user of the RadiForce series should assure that it is used in such an environment.
Immunity test
IEC60601 test level Compliance level
Electrostatic
±6kV contact
discharge (ESD)
±8kV air
EN61000-4-2
Electrical fast
±2kV
transient / burst
for power supply lines
EN61000-4-4
±1kV
for input/output lines
Surge
±1kV line(s) to line(s)
EN61000-4-5
±2kV line(s) to earth
Voltage dips, short
<5% U
interruptions and
U
) for 0.5 cycle
T
voltage variations on
40% U
power supply input
for 5 cycles
lines
70% U
EN61000-4-11
for 25 cycles
<5% U
U
) for 5sec
T
Power frequency
3A/m
(50/60Hz)
magnetic field
EN61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
The RadiForce series uses RF energy only for its internal function.
Therefore, its RF emission are very low and are not likely to cause any
interference in nearby electronic equipment.
The RadiForce series is suitable for use in all establishments, including domestic
establishments and those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic purposes.
±6kV contact
±8kV air
±2kV
for power supply lines
±1kV
for input/output lines
±1kV line(s) to line(s)
±2kV line(s) to earth
(>95% dip in
<5% U
T
T
U
) for 0.5 cycle
T
(60% dip in U
)
40% U
(60% dip in
T
T
T
U
) for 5 cycles
T
(30% dip in U
)
70% U
(30% dip in
T
T
T
U
) for 25 cycles
T
(>95% dip in
<5% U
T
T
U
) for 5sec
T
3A/m
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
(>95% dip in
Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the RadiForce series requires continued
operation during power mains interruptions, it
is recommended that the RadiForce series be
powered from an uninterruptible power supply or a
(>95% dip in
battery.
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.