2 | Advanced Adjustments/Settings
Note
• The gain value changes according to the color temperature.
• When the gain is changed, the color temperature setting changes to "User".
Reset
Resets any color adjustments for the currently selected mode back to the default settings.

2.2.2 RadiCS SelfQC

This product is equipped with an Integrated Front Sensor (calibration sensor). The
Integrated Front Sensor enables you to perform quality control (calibration and Grayscale
Check) on the monitor independently.
Note
• Calibration can be performed periodically by using the schedule function of RadiCS LE that is
attached to the monitor. For more information about the setup procedure, refer to the RadiCS LE
User's Manual (stored on the CD-ROM).
• RadiCS enables you to configure the schedule settings for Grayscale Check as well as for
calibration. For more information about the setup procedure, refer to the RadiCS User's Manual.
• If you want to perform high-level quality control in compliance with medical standards /
guidelines, use optional monitor quality control software kit "RadiCS UX2".
• To adjust measurement results of the integrated calibration sensor (Integrated Front Sensor) to
those of an external sensor that is sold separately, perform correlation between the Integrated
Front Sensor and the external sensor using RadiCS (optional accessory) / RadiCS LE (included).
Periodical correlation allows you to maintain the measurement result of the Integrated Front
Sensor at a level equivalent to that of the external sensor. For details on correlation, refer to the
RadiCS / RadiCS LE User's Manual.