Eizo RadiForce MX317W Installation Manual page 11

Color lcd monitor
Hide thumbs Also See for RadiForce MX317W:
Table of Contents

Advertisement

SelfCalibration
Execute SelfCalibration and view the latest results.
SelfCalibration - Execute
Executes SelfCalibration.
Attention
• If the power to the PC is turned off or the input signal switches while SelfCalibration is running,
the execution will be canceled.
• After SelfCalibration is complete, perform Grayscale Check at the same room temperature and
illuminance as those in the actual working environment.
Note
• The following items can be set by using RadiCS / RadiCS LE. For more information about the
setup procedure, refer to RadiCS / RadiCS LE User's Manual.
– Calibration target
– SelfCalibration schedule
– If you have configured the schedule using RadiCS / RadiCS LE, the next execution schedule
is displayed.
• The calibration target value can be confirmed by selecting the target mode in "CAL Switch Mode"
of the Setting menu.
SelfCalibration - Result
Settings: "DICOM" / "CAL" / "Patho"
The SelfCalibration execution results (such as Max. Error Rate, Last Time, and Error Code)
for the selected mode are displayed.
Attention
• The results can be checked when "QC History" is set to "On" (see
Settings - QC History
[} 13]).
Advanced Adjustments/Settings | 2
Settings - Warning
[} 13],
11