5    Service | Mattress; Replacing The Cpr Connector; Replacing The Air Cell Set - Invacare Softform Active 2 Rx Service Manual

Powered hybrid support surface
Hide thumbs Also See for Softform Active 2 Rx:
Table of Contents

Advertisement

Invacare® Softform Active® 2 Rx
5    Service | Mattress
5.1    Replacing the CPR Connector
Refers to spare parts kit SP1633237
1. Switch off the control unit and disconnect from mains
supply.
2. Open the zipper of the mattress and remove the top cover.
3. Remove the orange castellated foam insert.
4. Find the CPR connector at the mattress foot end. Unplug it at
the tube's extremities A.
5. Slide the defective connector off its sleeve and replace it
with the new one.
6. Plug the new connector onto the tube extremities.
7. Place the foam insert back on top of the mattress and close
the mattress cover's zipper.
5.2    Replacing the Air Cell Set
Refers to spare parts kit SP1627630 / SP1629718 / SP1629719
1. Switch off the control unit and disconnect from mains
supply.
2. Open the zipper of the mattress and remove the top cover.
3. Remove the orange castellated foam insert.
4. Find the CPR connector at the mattress foot end. Unplug it at
the tube's extremities (see 5.1 Replacing the CPR Connector,
page 17).
5. Detach the tubes connected to the T-valve A located on the
penultimate air cell.
6. Slide the T-valve through the strap's hole.
7. Thread the strap through the buckle B positioned on the
blue foam's lower side.
8. Discard the defective air cell set.
9. Repeat steps 5 to 7 for the five remaining T-valves.
10. Install the new air cell set and secure it with the provided
straps and buckles.
11. Place the foam layer back on top of the mattress and close
the mattress cover's zipper.
17
A
A
A
A
B
60129676-D

Advertisement

Table of Contents
loading

Table of Contents