Invacare® Softform Active® 2 Rx
4.6    Replacing the CPR Connector
Refers to spare parts kit SP1633238
NOTICE!
Risk of damaging the PCB by electro-static discharge
Handle PCBs only in Electrostatic Protected Areas
PH1 Philips Key
1. Switch off the control unit and disconnect from mains
supply.
2. Remove the bottom housing (see 4.1 Replacing the Front
Housing, page 6, step 4).
3. Unplug both silicone tubes A from the connector.
4. Discard the defective connector.
5. Reinstall the silicone tubes to the new connector and slide it
into the slit. Ensure no tube is kinked or clamped.
6. Reinstall the bottom housing.
4.7    Replacing the Tubing Set
Refers to spare parts kit SP1633112
NOTICE!
Risk of damaging the PCB by electro-static discharge
Handle PCBs only in Electrostatic Protected Areas
PH1 Philips Key
1. Switch off the control unit and disconnect from mains
supply.
2. Remove the bottom housing (see 4.1 Replacing the Front
Housing, page 6, step 4).
3. Unplug tubes A to G. Discard them.
4. Install the 320 mm long tube on one side of the T-valve and
the 40 mm long tube on the opposite side. Connect the third
(smaller diameter) tube both to the T-valve and to the
UIPCB.
5. Install remaining tubes to their original positions (see 4.5
Replacing the PCB, page 10, steps 14 - 17).
6. Reinstall the bottom housing.
13
A
B
C
D
A
A
E
F
G
60129676-D