2 Safety
2.1 General Safety Information
WARNING!
Risk of injury or damage to property
— The procedures in this manual must only be
performed by a qualified technician.
— Use only original accessories and spare parts.
— Do not handle this product or any available
optional equipment without first completely
reading and understanding these instructions
and any additional instructional material such
as user manuals, installation manuals or
instruction sheets supplied with this product o r
optional equipment.
— After each assembly, check that all fittings are
properly tightened and that all parts have the
correct function.
WARNING!
Risk of contamination
— Clean and disinfect the product before
servicing.
NOTICE!
Assembly of accessories might not be described i n
this service manual. Refer to the manual, d elivered
with the accessory.
— Additional manuals can be ordered from
Invacare. See addresses at the end of this
document.
NOTICE!
Replacement parts for this product family are only
available as a kit. Always use the complete new kit
when replacing a part.
— Spare parts kits can be ordered from Invacare.
Refer to your local Invacare website to access
the electronic spare parts catalogue (ESPC).
NOTICE!
Refer to the user manual of this product for
information on:
— Technical Data
— Product Components
— Labels
— Additional safety instructions
— Cleaning and disinfection instructions
NOTICE!
When components are replaced, it is n ecessary to
ensure their traceability by keeping the original
serial number in the control unit ( replacement of
spare parts with serial n umbers).
— Ensure the serial number traceability by either
printing a new serial label or transferring the
existing one, and placing it in the replacement
part.
The information contained in this document is s ubject
to change without notice.
4
2.2 Personal Safety Information
These safety instructions are intended to help avoid accidents
during work and must be observed under all circumstances.
All employees coming into contact with contaminated products
must regularly consult a company doctor. Work clothing and
personal protective equipment must be available in necessary
quantities and be in proper condition. Reliable hand and
surface disinfection must be ensured.
WARNING!
Risk of contamination
— Clean and disinfect the product before carrying
out repairs.
2.3 Electrical Safety Check
NOTICE!
Regular electrical checks according to EN 62353
(VDE0751) must be carried out in line with national
requirements.
Inspect all electronic components for damage to the casing or
cable, test functions and check for stray current.
2 Safety