Adjustment Procedure

The Agilent 83438A has only one adjustment procedure, total power. Periodic
adjustment is not required to maintain safety.
Option 022 instruments have an angled-fiber output. Be sure to use an angled-
fiber patchcord during testing. If you do not have an angled-fiber patchcord
available, you can purchase the required accessories from Agilent
Technologies. These include the Agilent 81000SI DIN connector interface and
the Agilent 81113PC DIN 4108 (angled) to Super-PC patchcord.
To avoid exposure to the laser path of a CLASS IIIb LASER
PRODUCT, do not open the A2A1 assembly. There are no
serviceable components inside. Do not damage the fiber-optic
cable that connects the CLASS IIIb pump laser, A2A3, to the
A2A1 assembly. Other laser paths outside of the A2A1
assembly do not require precautions to maintain safety.
Servicing

Adjustment Procedure

5-23