Agilent Technologies 8510XF Service Quick Reference Manual page 71

Network analyzer system
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

Self Test (1-14), Running Error Messages (15-41), Unratioed Power Test (42), LEDs/GPIB Addresses/Configuration Switch Settings
GPIB
System
Bus
69261
From
Sheet 1
9
69272
(In 8510XF for blanking only)
10 MHz REF
Trigger IN
NOTE: A16, A23, A22 are not used for
8510XF operation; that is, phase lock
none, except for some signal transfers
from LPTHLD, etc.
rr53c.cdr
GPIB
MOTHERBOARD
CAPACITORS
B1 FAN
08756-20073
FAN REPLACEMENT KIT
for 85102A
08756-60074
PC/CONTROLLER
GPIB
(VGA Type
Connector)
85101-60263
85101-60298
69272
85101-60300
7167
1
0950-3488
A15 LCD
A16 INVERTER
ASSEMBLY
BOARD
85101-60297
0950-3379
(Includes A16)
(A16 Only)
85101-60296
GPIB
8510XF System Level Troubleshooting Block Diagram (LCD) (Sheet 2 of 2)
SELF TESTS
1. A5 PROCESSOR EPROM (self test and service program
PROM using checksum)
2. A5 PROCESSOR RAM (test processor memory using data
complement technique)
3. A7 DATA BUS TEST (test I/O data bus)
4. A14 DISPLAY PROCESSOR
5. A14 DISPLAY RAM
6. A7 TIMER/CLOCK/RS – 232
7. A7 PUBLIC GPIB (Read/Write test of GPIB register)
8. A7 SYSTEM BUS (Read/Write test of system bus register)
9. INTERRUPT SYSTEM (test interrupts)
10. A5 MULTIPLIER (PERFORMS a complex multiply)
11. A7 DISC CONTROLLER
12. A6 NON-VOLATILE MEMORY TEST (read/write test)
13. IF DETECTOR DATA (read/write test data and address lines)
14. KEYBOARD (read 85101C keyboard for a stuck key)
RUNNING ERROR MESSAGES
15. FAILED PRETUNE/LOST PRETUNE
16. NO IF FOUND
17. PHASE LOCK FAILURE
18. PHASE LOCK LOST
19. VTO OVER-RANGE
20. ADC CAL FAILED
21. IF CAL FAILED – REFERENCE;
IF CAL FAILED – TEST;
IF CAL FAILED – BOTH (channels)
22. ADC NOT RESPONDING
23. IF OVERLOAD (External level of 8360 can cause this.)
24. AUTORANGE CAL FAILED
25. SOURCE SLEEP SYNC ERROR
26. SWEEP TIME TOO FAST
27. SOURCE (GPIB) SYNTAX ERROR
28. TEST SET (GPIB) SYNTAX ERROR
29. SYSTEM BUS ADDRESS ERROR
30. SYSTEM BUS SRQ ERROR
31. UNABLE TO LOCK TO EXT 10 MHz REF
32. ERROR: SYSTEM KEY NOT INSTALLED; CAUSED BY SECURITY
KEY MISSING, BAD I/O, BAD A8 CONNECTOR, BAD MOTHER-BOARD.
CAUTION: OPTIONAL FUNCTION: NOT INSTALLED
CAUTION: TIME DOMAIN KEY NOT INSTALLED
33. DISC HARDWARE PROBLEM
34. DISC COMMUNICATION ERROR
35. INITIALIZATION FAILED
36. DISC MEDIA WEARING OUT - REPLACE SOON
8360 – SERIES SOURCES RUNNING ERROR MESSAGES
37. FAILURE – FAULT INDICATOR ON
38. FAILURE – RF UNLOCKED
39. FAILURE – OVERMODULATION
40. FAILURE – SELF TEST FAILURE
41. INVALID SETUP – CHECK SYSTEM BUS CONFIGURATION
OTHER TROUBLESHOOTING HINTS
42. Perform the unratioed power test for a1, a2, b1, and b2. Refer to
'System Maintenance, Troubleshooting' in the 8510XF Operating and
Service Manual, E7350-90001.
43. The A14 GSP and the A15 display must be working in order for a normal
display to be on the LCD.
44. If self test 11 passes, try to initialize a blank disc to further test the A2
disc drive.
45. A18 (85102) ANALOG TO DIGITAL CONVERTER TEST
When you first turn the instrument ON, if the noise on the trace is
greater than 1 dB, then the A18 board may be faulty.
If you cycle the AC power (Off, then On), the noise may disappear.
If it does, there is probably a malfunction with the A18 A/D board
although no apparent symptoms are present. There also may be no
Running Error Messages, Self Test Failures or Service Program test
failures.
If this trace noise disappears when the POWER is cycled, the test has
FAILED and the A18 board should be REPLACED.