NOTE
In Case of Difficulty
3. To set the frequency using the analyzer front panel, press
[CENTER] [MENU] {SINGLE POINT}
Enter the start frequency of the source.
4. Measure RF and LO frequencies with the counter, and record these
values on the test record at the end of this section.
5. From the analyzer front panel, enter the frequency per the tables at the
end of this section.
Be sure to connect RF and LO output to the 500 MHz to 26.5 GHz input on
the frequency counter. Also set the input switch to the 500 MHz to 26.5 GHz
position.
6. Measure the frequency with the counter, then record the value on the test
record at the end of this section.
If the measured values do not meet the specifications listed on the test
record, refer to your source manual for adjustment and troubleshooting
instructions.
Performance Verification
CW Frequency Accuracy Test
.
8510XF Service Quick Reference Guide 4-13