Agilent Technologies 8510XF Service Quick Reference Manual page 32

Network analyzer system
Hide thumbs Also See for 8510XF:
Table of Contents

Advertisement

Table 4-2
Hardware
Selection (<50 GHz Range)
Network Analyzer
8510C -- Vector Network Analyzer
Test Set
E7352A -- 1.0 mm Subsystem (2-50 GHz)
E7352A005 -- 1.0 mm Subsystem (45MHz-50GHz)
Source
8365xXF -- Synthesizer (45 MHz-50 GHz)
1
Calibration Kit
85059A -- 1.0 mm (45 MHz-50 GHz) [Use w/
E7342A/52A]
Calibration
BL -- Broadband Load Cal
Technique
Test Port Cables
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
Verification Kit
85059AOV -- 1.0 mm (45 MHz-50 GHz)
1. The 85059A --1.0 mm (45MHz-50GHz) [Use w/ 8517A/B] selection is not used for the 8510XF Performance Verification.
Table 4-3
Hardware
Selection (<50 GHz Range)
Network Analyzer
8510C -- Vector Network Analyzer
Test Set
E7342A -- 1.0 mm Subsystem (2-50 GHz)
E7342A005 -- 1.0 mm Subsystem (45 MHz-50GHz)
Source
8365xXF -- Synthesizer (45 MHz-50 GHz)
1
Calibration Kit
85059A -- 1.0 mm (45 MHz-50 GHz) [Use w/
E7342A/52A]
Calibration
BL -- Broadband Load Cal
Technique
Test Port Cables
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
Verification Kit
85059AOV -- 1.0 mm (45 MHz-50 GHz)
1. The 85059A --1.0 mm (45MHz-50GHz) [Use w/ 8517A/B] selection is not used for the 8510XF Performance Verification.
Frequency ranges
For most of the menu items in the software, there are two or more selections
that pertain to the 8510XF. These menu selections are differentiated by:
A. The frequency range of the millimeter wave subsystem
E7352A: 110 GHz subsystem
E7342A: 85 GHz subsystem
Option 005: 45 MHz to 2 GHz
B. The portion of the subsystem's frequency range that is being verified (the
ranges above and below 50 GHz are verified separately).
The menu selections are broken down in Tables 4-2 and 4-3, below.
System hardware configuration (E7352A -- 110 GHz System)
System hardware configuration (E7342A -- 85 GHz System)
Performance Verification
Selection (>50 GHz Range)
8510C -- Vector Network Analyzer
E7352A110 -- 1.0 mm (50 GHz-110 GHz)
836xxXF -- Multiplied Synth. (above 50 GHz)
85059A110 -- 1.0 mm (50 GHz-110 GHz)
OS -- Offset Short Cal
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
85059AOV110 -- 1.0 mm (above 50 GHz)
Selection (>50 GHz Range)
8510C -- Vector Network Analyzer
E7342A85 -- 1.0 mm (50 GHz-85 GHz)
836xxXF --
Synth. (above 50 GHz)
85059A85 -- 1.0 mm (50 GHz-85 GHz)
OS -- Offset Short Cal
DIRECTXF -- for 8510XF verification (1.0 mm-1.0 mm)
85059AOV110 -- 1.0 mm (above 50 GHz)
8510XF Service Quick Reference Guide 4-3