9. Verify
Cal Set for 45 MHz Measurement (If Required)
10. Select
and follow the program prompts.
Done
11. Select
to measure the verification standards. Follow the
Measure Data
program prompts.
12. When verification has been completed view the data, then print or save
(if required). If needed the verification standard can be remeasured by
selecting:
Repeat Measure
13. Select
to return to the Standard Selection Menu. Repeat
Select Standard
steps 7
through
12
above for the other verification standard (refer to
Table 4-6 on page
4-9).
14. Select
until the Main Menu appears.
Prior Menu
15. Select
to exit the program, then close the program window to
Quit Program
finish.
Performance Verification
Verification Procedures
is set to
.
8510XF Service Quick Reference Guide 4-11
.
No