Table 2-1
Model
83620B/21B
83650A/B
83651A/B
83621A
83631A
83620A
83623A
83624A
83630A
83640A
1. Fully compatible at time of shipment.
2. Can be used for 8510XF LO source when upgraded and includes Options 004 and 008.
3. Includes installation.
4. 83623L, 30L, 40L and 50L, with Option 004 and 008, will also work as LO source, but do not have
modulation capability (83623L maximum power at highest frequency is + 15 dBm).

Upgrading a Source for 8510XF

All 8360 series sources are compatible, but only the 83621A/B or 83651A/B
are recommended in order to insure full functionality of the 8510XF.
Table 2-1
shows the upgrade options required.
83651/21A and 83650/20A/B will also work for upgraded 8510XF systems
as long as it has test port flatness correction modification or capability, rear
panel output (Option 004), and 1 Hz frequency resolution (Option 008)
installed.
Source Upgrades
Maximum Power at
Highest Frequency
+ 13 dBm
+ 2.5 dBm
+ 2.5 dBm
2
+13 dBm
2
+ 10 dBm
2
+ 13 dBm
2 4
+ 17 dBm
2
+ 20 dBm
2
+ 10 dBm
2
+ 26 dBm
Serial Prefix
All
< 3103A
3103A
3104A to 3111A
≥ 3112A
≤ 3103A
3104A to 3111A
≥ 3112A
≥ 3145A
8510XF Service Quick Reference Guide 2-1
Required for Test Port
Flatness Correction
1
No modification required.
3
83601A upgrade kit.
08360-60167 firmware kit
08360-60201 firmware kit
1
No modification required.
08360-60167 firmware kit
08360-60201 firmware kit
1
No modification required.
1
No modification required.