Connecting the System

WARNING: All connections to the OPTIS Integrated Next except the TSC in
Bluetooth
mode must be made through the System Connector Panel. Making
connections directly to internal components of the system may bypass isolation
features and compromise patient safety.
WARNING: If the external monitor is being used in the patient vicinity, it must
use an isolated power source or it may compromise electrical isolation and
cause patient injury.
The system cabinet and monitors are normally always connected to power.

Powering On

WARNING: Never use a converter adapter to connect the AC plug to an
ungrounded wall outlet. Doing so may result in electric shock to the patient
or operator and damage to equipment.
For detailed information on electrical requirements, refer to "Making Proper
Electrical Connections" on page 10.
Make sure the power cords are connected to the system cabinet and to the
1.
control room monitor, and that they are plugged into grounded electrical
outlets.
2.
Confirm that all monitors are powered by observing that the appropriate power
indicators are on.
3.
Turn ON system power by pressing the Standby button located at the
upper-right side of the front of the cabinet. The startup screen appears.
The first time the software runs, select a language (if prompted). If necessary,
4.
refer to the installed software User Manual to change this selection.
The first time the software runs, the End User License Agreement (EULA)
5.
displays. Read and agree to the EULA to proceed.
6.
Adjust monitor brightness and contrast as necessary.
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A