Maintenance - Abbott OPTIS 1014933 Operation And Maintenance Manual

Integrated next imaging system
Table of Contents

Advertisement

State: Released Date: 2021.03.01 22:19 GMT Effectivity: Upon Release
Cleaning and Maintenance

Maintenance

Contacting Abbott Technical Service
CAUTION: Only a qualified Abbott Technical Service representative can perform
maintenance or service components of the system. Any attempt to open the
system components by anyone other than a qualified Abbott Technical Service
representative will void the warranty.
Service can be contacted at:
Email: OCTservice@abbott.com
Phone (Outside US): +1 651 490 4410
Maintenance Procedures
Maintenance of the system consists of:
Optical Connection Cleaning Procedure
The optical connection between the DOC and the catheter should be cleaned
whenever a loss of image quality occurs. Image quality should also be checked
every 3 months, and the connection should be cleaned if there is a loss of
quality.
CAUTION: This procedure should not be performed during a patient case on a
sterile Dragonfly Imaging Catheter.
Note:
1.
2.
30
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
Cleaning the optical connection in the DOC and in the Dragonfly™ Imaging
Catheter
Replacing the optical adapter in the DOC
Inspecting exposed cable connections
Transferring log files
Identifying the installed software version
Before beginning this procedure, ensure that the Abbott Medical optical
fiber connector cleaner is available.
Do not touch any of the optical connectors or the end of the optical fiber
connector cleaner as this may damage them.
If a catheter is connected to the DOC, press the Unload button on the DOC
and wait until the "lock" LED stops flashing. When the "lock" LED is off, remove
the catheter.
Click the Menu button and select the Setup option. The Setup dialog box
displays.