Cleaning and Maintenance

CAUTION: Do not perform cleaning or maintenance on the system during a
patient case or in the patient environment.
CAUTION: The OPTIS™ Integrated Next accessories should be appropriately
classified as a biohazard and disposed of in compliance with applicable facility
procedures, local and country laws and regulations. Return the OPTIS
Integrated Next to Abbott Medical at the end of its operating life.

Cleaning

The following OPTIS Integrated Next items require cleaning:
DOC and DOC optical cable
TSC
DOC holster and cables
The DOC, DOC holster, and TSC should be cleaned following the facility's
standard cleaning schedule, or at least every 30 days under normal use.
1.
Turn off (or Standby) all system components with accessible power controls
and unplug the power cables.
Clean system surfaces and the keyboard with a dry cloth.
2.
3.
Clean the LCD surface of the monitor with a lint-free, non-abrasive cloth.
CAUTION: Do not clean the LCD surface with detergents or other cleaning
solutions.
Clean the TSC, DOC, DOC holster, and DOC optical cable with a disinfectant
4.
wipe (1:10 Hypochlorite solution) or Cidex
cloth. Be particularly careful not to stress or sharply bend the DOC optical
cable.
Note:
Though enclosed in a bag during use, the DOC is the system component
most exposed to dirt, fluids, and debris.
The TSC and DOC are not waterproof. Be careful not to use excessive
moisture when cleaning these devices. Wipe only. Do not soak or spray.
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
(Glutaraldehyde 3.4%) and a soft
29