State: Released Date: 2021.03.01 22:19 GMT Effectivity: Upon Release
System Setup

Shutting Down

CAUTION: Do not unplug from AC power or turn off main power until the
shutdown is complete, the screens turn black, and the green monitor LEDs turn
amber. Disconnecting from AC power before the shutdown is complete may
damage the system.
Note: Use the Shutdown button at the top of the screen to shut down the
system. Use of the On / Standby key is not recommended to shut down
the system.
1.
2.
3.
4.
28
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
Click the Shutdown button located at the top of the screen. The system
shutdown menu appears.
Click Yes to begin the system shutdown, or No to continue using the system.
If you choose Yes, the computer begins the system shutdown. After
15 seconds, the screens turn black, the green monitor LEDs turn
amber, and the system enters Standby mode.
After the screens turn black and the green monitor LEDs turn amber, press the
main power switch at the base of the cart to turn off system power.
If necessary, disconnect the power cord from AC power.