Abbott OPTIS 1014933 Operation And Maintenance Manual page 37

Integrated next imaging system
Table of Contents

Advertisement

State: Released Date: 2021.03.01 22:19 GMT Effectivity: Upon Release
Optical Adapter Replacement Procedure
The optical adapter should be replaced every 200 cycles or 1 year (whichever
occurs first), or if cleaning does not improve image quality. Contact Abbott
Service to purchase optical adapters.
CAUTION: This procedure should not be performed during a patient case on a
sterile Dragonfly Imaging Catheter.
Note:
Before beginning this procedure, ensure that a replacement Abbott
Medical optical adapter and supplied pliers are available.
Do not touch any of the optical connectors, as this may damage them.
If a catheter is connected to the DOC, press the Unload button on the DOC
1.
and wait until the "lock" LED stops flashing. When the "lock" LED is off,
remove the catheter.
Click the Menu button and select the Setup option. The Setup dialog box
2.
displays.
3.
Click Service. The Service menu displays.
In the DOC Service section of the Service menu, click Enter. The DOC
4.
optical carriage moves all the way to the front and locks into position for one
minute. Rotation of the DOC rotary motor stops and the laser source turns
off.
Use pliers to grip either short surface of the adapter as shown (figure
5.
below). Ensure that the pliers seat over the lip on the end of the adapter.
Figure 10. Proper Gripping of Adapter for Removal
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
Cleaning and Maintenance
33