State: Released Date: 2021.03.01 22:19 GMT Effectivity: Upon Release
Table 6. Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The OPTIS Integrated Next is intended for use in the electromagnetic environment specified below. The customer or user of the OPTIS Integrated
Next should ensure that it is used in such an environment.
Emissions Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations / flicker emissions
IEC-61000-3-3
Note: The emission characteristics of this equipment make it suitable for use in industrial areas and hospitals (CISPR 11
class A). If it is used in a residential environment (for which CISPR 11 class B is normally required), this equipment might
not offer adequate protection to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
Table 7. Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user of the device should ensure that it is
used in such an environment.
Immunity Test
IEC 60601 Test Level
Electrostatic discharge
±8 kV contact
(ESD)
±15 kV air
EN 61000-4-2
(IEC 61000-4-2)
Electrical fast
±2 kV for power supply
transient / burst
lines
EN 61000-4-4
±1 kV for input / output
(IEC 61000-4-4)
lines
Surge
±1 kV differential mode
EN 61000-4-5
±2 kV common mode
(IEC 61000-4-5)
Voltage dips, short
0% U
; 0.5 cycle
T
interruptions, and
At 0°, 45°, 90°, 135°,
voltage variation on
180°, 225°, 270° and
power supply input
315°
lines.
0% U
; 1 cycle
T
IEC 61000-4-11
and
70% U
; 25/30 cycles
T
Single phase: at 0°
0% U
; for 5 sec @ 60 Hz
T
(300 cycles)
0% U
; 250/300 cycles
T

Electromagnetic Emissions

Compliance
Electromagnetic Environment Guidance
Group 1
The OPTIS Integrated Next uses RF energy only for its internal function.
Therefore, its RF emissions are very low and not likely to cause any
interference with nearby electronic equipment.
Class A
The emission characteristics of the OPTIS Integrated Next are suitable for
use in industrial areas and hospitals (CISPR 11 class A). System should not
be used in a residential (domestic) environment (CISPR 11 class B is
required in support of domestic use).
Class A
Complies

Electromagnetic Immunity

Compliance Level
±8 kV contact
±15 kV air
±2 kV
±1 kV
±1 kV
±2 kV
100% dropout in
VNOM for 0.5 cycle at
listed phase angles
100% dropout in
VNOM for 1 cycle at 0°
30% dropout in VNOM
for 25/30 cycles at 0°
100% dropout in
VNOM for 5 sec
100% interrupt in
VNOM for 250/300
cycles
OPTIS™ Integrated Next Operation and Maintenance Manual – 600173796 Rev. A
System Specifications
Electromagnetic Environment - Guidance
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should be
at least 30%.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial or
hospital environment.
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the device requires
continued operation during power mains interruptions, it is
recommended that the device be powered from an
uninterruptible power supply or a battery.
41