Return Loss Test - Agilent Technologies 81632B User Manual

Power sensor modules
Table of Contents

Advertisement

42
6 Move to <AvgTime>, press Enter, move to <1 s>, and press Enter.
7 Move to <Range mode>, press Enter, move to <Auto>, and press
Enter, and press [Close].
8 Press the [Parameter] softkey, move to [TotalTime], press Enter, set
the total time to 00:05:00, which is 5 minutes, and press Enter.
9 Press the [Measure] softkey to start the measurement.
10 After the stability application has finished, press the [Analysis] softkey ,
press the [more] softkey , and note [∆ P ] as the noise value in the test record.

Return Loss Test

1 Make sure that all connectors are clean.
2 Set up the equipment as shown in Figure 6.
3 Press [Preset] on the mainframe.
Figure 6
Return Loss Measurement Setup
N O TE
To ensure traceability, use the 81610CC Reference Cable for
calibration measurements. Do not use the 81610CC Reference Cable
for measurements on a Device Under Test. Instead, use a measurement
patchcord.
Agilent 81632B/3B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power
Sensor Module, First Edition
Mainframe 8163B or 8164B w/
81657A and 8163xA/B (DUT)
8163 B
81613A 8163xB
81635A
in
out
81610CC