Agilent 81632B/3B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, First Edition
26 Decrease the attenuation of Atty2 by 10 dB in order to be on the
upper limit of the -50 dBm range.

-10 dBm Range

27 Disable Atty1.
28 Zero both Power Meters. On the 8163B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
29 Enable Atty1.
30 Switch one range up to the 0 dBm range and note the power
readings (#6).
31 Switch down to the previous range (-10 dBm) and note the values
again (#7).
32 Increase the attenuation of Atty1 by 10 dB and note the results in
the test record (#8).
33 On the DUT switch one range down to the -20 dBm range.
34 Decrease the attenuation of Atty2 by 10 dB in order to be on the
upper limit of the -50 dBm range.
-20 dBm Range
35 Disable Atty1.
36 Zero both Power Meters. On the 8163B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
37 Enable Atty1.
38 Switch one range up to the -10 dBm range and note the power
readings (#9).
39 Switch down to the previous range (-20 dBm) and note the values
again (#10)
40 Increase the attenuation of Atty1 by 10 dB and note the results in
the test record (#11).
41 On the DUT switch one range down to the -30dBm range
42 Decrease the attenuation of Atty2 by 10 dB in order to be on the
upper limit of the -50 dBm range.
-30 dBm Range
43 Disable Atty1.