Table 4
Fast Optical Power Sensor Module Specifications
Agilent 81636B
Agilent 81637B
included
Analog Output
75 mm × 32 mm × 335 mm (2.8" × 1.3" × 13.2")
Dimensions (H x W x D)
0.5 kg
Weight
2 years
Recalibration Period
+10 ° C to +40 ° C
Operating Temperature
Non-condensing
Humidity
20 minutes
Warm-up time
1 Reference Conditions
• Power level 10 µ W (
¥
20 dBm), continuous wave (CW)
• Fiber 50 µ m graded-index, NA=0.2
• Ambient temperature 23 ° C ± 5 ° C
• On day of calibration (add ± 0.3% for ageing over one year; add ± 0.6% over two years)
• Spectral width of source < 10 nm (FWHM)
• Wavelength setting at Power Meter must correspond to source wavelength ± 0.4 nm
2 Operating Conditions:
• Fiber ≤ 50 µ m, NA ≤ 0.2
• Only Agilent 81632B, Agilent 81635A and Agilent 81636B: For fiber 62.5 µ m graded-index (NA=0.24):
add ± 2 %
• Within one year of calibration, add 0.3% for second year
• Add ±1% for Biconic connector
• Operating temperature as specified, humidity: non-condensing
3 All states of polarization at constant wavelength (1550 nm ± 30 nm) and constant power, straight connector, T=23 ° C ± 5 ° C
For angled connector (8 ° ) add ± 0.01 dB typ.
4 Conditions:
Wavelength 1550 nm ± 30 nm, fixed state of polarization, constant power, Temperature 23 ° C ± 5 ° C
Linewidth of source ≥ 100 MHz, angled connector 8 ° .
5 At constant temperature ∆ T = ± 1 ° C
6 Averaging time 1 s, T = 23 ° C ± 5 ° C, observation time 300 s. Wavelength range 1260 - 1630 nm.
25
Agilent 81632B/3B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, First Edition