Zero both Power Meters. Move to <Zero all> and press Enter.
10 Enable both attenuators.
11 Adjust the attenuation of Atty2 in order to achieve -47.2 dBm on the
reference Power Meter.

+10 dBm Range

12 Note both power readings as the first value in the test record, which
is given at the end of the test descriptions (#1).
N O TE
Always include the three digits after the decimal point when you
note a power reading.
13 Increase the attenuation of Atty1 until the power reading of the
DUT shows about +2.8 dBm.
14 Note the InRange-values in the test record (#2)
15 Perform the following sub-procedure for the DUT, 8163xA:
a Press [Menu], move to <Range mode>, move to <Manual>, and press
Enter.
b Move to <Range>, press Enter, move to <0 dBm>, and press Enter.
16 If necessary, adjust the attenuation of Atty2 in order to be on the
upper limit of the -50 dBm range (i.e. -47.2 dBm).
0 dBm Range
17 Disable Atty1.
18 Zero both Power Meters. On the 8163B with two installed power
meters, press [Menu], move to <Zero all>, and press Enter.
19 Enable Atty1.
20 Switch one range up to the +10dBm range.
21 Note both power readings (#3).
22 Switch down to the previous range (0dBm) and note the values
again (#4).
23 Increase Atty1 by 10 dB and note the results in the test record (#5).
24 Move to the channel of the DUT, 8163xA, press [Menu], move to
<Range>, press Enter, move to <-10 dBm>, and press Enter.
25 If necessary, adjust the attenuation of Atty1 in order to be on the
upper limit of the range (i.e. -x7.y dBm).
Agilent 81632B/3B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power
Sensor Module, First Edition