Performance Test for the Agilent 81636B
Page 3 of 3
Model Agilent 81636B Sensor Module
Test No. Test Description
I
Accuracy Test
measured at _______ nm
(1310nm)
measured at _______ nm
(1550nm)
II
Linearity Test
Range
+10
+10
0
0
-10
-10
-20*
-20
-30
-30
-40
-40
-50
-50
III
Noise Test
IV
Return Loss Test
Agilent 81632B/3B/4B/6B/7B Power Sensor Modules and Agilent 81635A Dual Power Sensor Module, First Edition
Output Power
Output Power
P
[dBm] P
[dBm]
DUT
DUT
+9
________
+3
________
+3
________
-7
________
-7
________
-17
________
-17*
Reference
-27
________
-27
________
-37
________
-37
________
-47
________
-47
________
-57
________
40 dB <
Report No.________
Min.
Spec.
Result
[µW]
9.64 µW
_______
9.64 µW
_______
For Calculations you may want to use the appropriate sheet
Loss [%]
_______
_______
_______
_______
_______
_______
0.0
_______
_______
_______
_______
_______
_______
_______
[pW]
_______
< 20 pW
[dB]
_______
Date ________________
Max.
Measurement
Spec.
Uncertainty
10.36 µW
10.36 µW
<± 0.92 %
<± 0.92 %
<± 0.92 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.46 %
<± 0.47 %
<± 0.47 %
<± 0.56 %
<± 0.56 %
<± 1.46 %