N5500A With Option 001; N5500A With Option 201 - Keysight Technologies Agilent N5500A Instruction Manual

Phase noise test set
Table of Contents

Advertisement

N5500A with Option 001

Option 001 adds an input attenuator, input switching, an AM detector, and a
microwave phase detector to the test set. The input attenuation ranges from 0
to 35 dB in 5 dB steps.
The RF ANALYZER output can view either an input signal of up to 26.5 GHz or
baseband noise. The SIGNAL input can be switched to AM detector, RF phase
detectors, microwave phase detector, or an external microwave
downconverter. The IF signal from the downconverter is switched to the RF
phase detectors.

N5500A with Option 201

The N5500A Option 201 adds a microwave phase detector to extend the carrier
frequency range to 26.5 GHz. The microwave phase detector frequency ranges
from 1.2 GHz to 26.5 GHz. The microwave phase detector input power at the
µW SIGNAL input ranges from 0 to +5 dBm and +7 to +10 dBm at the
REF INPUT connections.
Agilent N5500A Hardware Reference