Chapter 6, Memory Cards/Mass Storage

Default Mass Storage Locations

Default Mass Storage Locations
Built-in IBASIC Controller
Save/Recall Registers
340
The default mass storage location for the built-in IBASIC Controller is the front
panel memory card slot (mass storage volume specifier ":INTERNAL,4") after
any of the following conditions:
power-up
initializing RAM with the SERVICE screen's RAM Initialize function
The mass storage location for the built-in IBASIC Controller can be changed
using the MASS STORAGE IS command. Refer to the HP Instrument BASIC
Users Handbook for further information on the MASS STORAGE IS command.
The default mass storage location for the Save/Recall registers is the Test Set's
internal RAM (no mass storage volume specifier) after any of the following
conditions:
power-up
initializing RAM with the SERVICE screen's RAM Initialize function
resetting the Test Set using the front-panel PRESET key
resetting the Test Set using the *RST HP-IB Common Command
The mass storage location for Save/Recall registers can be changed using the
field in the I/O CONFIGURE screen. The default mass storage
Save/Recall
volume specifiers for the Save/Recall register mass storage locations are as
follows:
Internal selection - (no mass storage volume specifier, registers are saved to allocated
RAM space)
Card selection (not changeable) - ":INTERNAL,4"
RAM selection (not changeable) - ":MEMORY,0,0"
Disk selection - the External Disk Specification field in the TESTS
(External Devices) screen.