HP 8920A Programmer's Manual page 61

Rf communications test set
Hide thumbs Also See for 8920A:
Table of Contents

Advertisement

10
COM /Io_names/ INTEGER Inst_addr,Bus_addr
20
CLEAR SCREEN
30
Inst_addr=714
40
Bus_addr=7
50
CLEAR Inst_addr
60
OUTPUT Inst_addr;"TRIG:ABORT"
70
OUTPUT Inst_addr;"*RST"
80
OUTPUT Inst_addr;"DISP RFAN"
90
!
100 ! Execute a call to the Measure function with a request to measure RF
110 ! power.
The time out value is specified as 10 seconds. The value
120 ! returned by the function is assigned to the variable Measure_result.
130 !
140 Measure_result=FNMeasure("MEAS:RFR:POW?",10)
150 !
160 ! Check the result of the function call.
170 !
180 IF Measure_result=9.E+99 THEN
190
PRINT "Measurement failed."
200 ELSE
210
PRINT "Power = ";Measure_result
220 END IF
230 END
240 !***********************************************************
250 ! Recommended Technique:
260 !***********************************************************
270 DEF FNMeasure(Query_command$,Time_out_value)
280
COM /Io_names/ INTEGER Inst_addr,Bus_addr
290
DISABLE
300
ON TIMEOUT Bus_addr,Time_out_value RECOVER Timed_out
310
OUTPUT Inst_addr;"TRIG:MODE:RETR SING;:TRIG:IMM"
320
OUTPUT Inst_addr;Query_command$
330
ENTER Inst_addr;Result
340
OUTPUT Inst_addr;"TRIG:MODE:RETR REP"
350
ENABLE
360
RETURN Result
370 Timed_out:!
380
ON TIMEOUT Bus_addr,Time_out_value GOTO Cannot_recover
390
CLEAR Inst_addr
400
OUTPUT Inst_addr;"TRIG:ABORT;MODE:RETR REP"
410
ENABLE
420
RETURN 9.E+99
430 Cannot_recover:!
440
DISP "Cannot regain control of Test Set."
450
STOP
460 FNEND
Chapter 2, Methods For Reading Measurement Results
HP BASIC 'ON TIMEOUT' Example Program
61

Advertisement

Table of Contents
loading