Table 2
RX TEST Screen and TX TEST Screen Priority Field Preset Values
2. When entering the TX TEST screen,
a. The AF Anl In field, the De-Emphasis field, the Detector field and the
AF Analyzer Measurement field (measurement displayed in upper, right portion
of CRT display) are,
•
set to their preset values upon entering the screen for the first time since
power-up, OR
•
set to their preset values if the PRESET key is selected, OR
•
set to the last setting made while in the screen
b. The AF Analyzer AF Anl In, De-Emphasis and Detector fields are,
•
set to their preset values whenever entering the screen, OR
•
set to their preset values if the PRESET key is selected
RX TEST
Priority
Screen Preset
Field
Value
−80 dBm
RF Gen
Amplitude
AFGen1 To
FM
AF Anl In
Audio In
Detector
RMS
De-emphasis
Off
AF Analyzer
SINAD
Measurement
Field Hidden
TX TEST
On RX TEST
Screen Preset
Screen
Value
No
Off
No
Audio Out
Yes
FM Demod
Pk ± Max
Yes
750 µs
Yes
No
Audio Freq
Chapter 1, Using HP-IB
Getting Started
Field
Hidden On
TX TEST
Screen
Yes
Yes
No
No
No
No
37