Reading The Received Data Fields
from the mobile station on the reverse control channel or reverse voice channel or
the raw data message bits displayed when a decoding error occurs, the control
program queries one, some, or all of the six received data fields. The information
in each field is returned exactly as displayed on the CRT. The information is
returned to the control program as a quoted string ("This is an example.").
The received data fields are read only data fields.
The :RCDD1? through :RCDD6? query commands are used to read the contents
of the six received data fields.
Syntax
:RCDD<1-6>?
Example
OUTPUT 714;"CALLP:RCDD1?"
ENTER 714;Rcv_data$
Setting the Display Field to Meas
When the
field is set to
Display
screen is used to display modulation quality measurements made on the
CONTROL
mobile station's RF carrier while on a voice channel.
Four characteristics of the RF carrier are measured: TX Freq Error, TX Power,
FM Deviation, and AF Frequency. The
connected state (that is, the
Figure 36, "CALL CONTROL Screen with Meas Selected," on page 472
layout of the
CALL CONTROL
Chapter 8, Programming The Call Processing Subsystem
Programming The CALL CONTROL Screen
To read the decoded data messages received
the top right-hand portion of the
Meas
information is only available in the
Meas
annunciator is lit).
Connect
screen when
is selected.
Meas
CALL
shows the
471