Chapter 3, HP-IB Command Guidelines
Guidelines for Operation
Command Punctuation
NOTE:
Programming Language Considerations. The punctuation rules for the Test Set's
HP-IB commands conform to the IEEE 488.2 standard. It is possible that some
programming languages used on external controllers may not accept some of the
punctuation requirements. It is therefore necessary that the equivalent form of the correct
punctuation, as defined by the language, be used for HP-IB operation. Improper
punctuation will results in HP-IB Error: -102 Syntax Error.
72
Using Quotes for String Entries
Quotation marks ' and " are used to select a non-numeric field setting. The value is
entered into the command line as a quoted alphanumeric string.
Quotes are used with all Underlined (toggling) and One-of-many (menu choice)
fields. (See "Changing A Field's Setting" in chapter 1 of the HP 8920A User's
Guide for field type descriptions.)
For example, to set the RF Generator's
Dupl would be entered into the command string.
RFG:OUTP 'Dupl'
or
RFG:OUTP "Dupl"
Using Spaces
When changing a field's setting, a space must always precede the setting value in
the command string, regardless of the field type (command<space>value).
RFG:FREQ<space>850MHZ
RFG:ATT<space>'OFF'
field to
Output Port
(duplex), the
Dupl

Advertisement