Chapter 6:
Troubleshooting and Diagnostics
RESPIRONICS"
e
6.6
SmartCard Error Code Chart
Error Code
Problem
Corrective Action
À
1
Read Fail - Card orientation is wrong.
. Re-enter the SmartCard. Iferror still persists, replace SmartCard.
2
Invalid card type - orientation is wrong or wrong
Re-enter the SmartCard. Iferror still persists, replace SmartCard.
card.
3
Invalid Card Format - Information on SmartCard is
Place the SmartCard in the Reader/ Writer and re-enter the informa-
incorrectly entered.
tion back onto the SmartCard.
4
Unit in Dealer Mode.
Remove power from the Respironics CPAP System. Reconnect
power to unit without pressing any keys.
5
Unit in Calibration Mode.
Remove power from the Respironics CPAP System. Reconnect
power to unit without pressing any keys.
6
Invalid Card ID Block. The checksum in the
Remove and / or replace SmartCard.
SmartCard ID block is invalid.
10
Prescription Block Checksum Failure.
Use the SmartCard Wizard in Encore 3.0 or higher to format the
SmartCard.
101
Prescription Version Error.
Use the SmartCard Wizard in Encore 3.0 or higher to format the
SmartCard.
102
Invalid Unit. The unit is not capable of the specified
Use the SmartCard Wizard in Encore 3.0 or higher to format the
prescription.
SmartCard.
103
Prescription Format Error. Software cannot under
Use the SmartCard Wizard in Encore 3.0 or higher to format the
C
stand the format of the prescription.
SmartCard.
)
1011170
REMstar Pro CPAP System
6-17