Chapter 6:
Troubleshooting and Diagnostics
RESPIRONCS
Re
ANETO
ος
m
Troubleshooting Table (Continued)
( Symptom
Cause
Verification
Corrective Action
3
Pressure Offset -
The delivered pressure is higher or | Perform power up process.
* Check tubing for kinks.
Static pressure fluctuates from
・1.0 cm to+2.0 cm HLO.
lower than the set value by more
than -1.0 cm to +2.0 cm HO.
+ Replace the Main PCA.
No Stored Therapy Data
SmartCard was not installed in unit | No therapy data after usage.
* Depress the pressure on / off
prior to use.
button before disconnecting the
Main AC power.
Nota Respironics SmartCard.
Card not formatted or faulty
SmartCard.
X
A
1011170
REMstar Pro CPAP System
6-10