Error Code Chart (Continued)
Chapter 6:
Troubleshooting and Diagnostics
へ
(Error Code
Problem
Corrective Action
9
Motor Overspeed.
Replace blower assembly. If error persists, replace Main PCA.
10
Motor Underspeed - The speed of the motor was determined
Replace blower assembly. If error persists, replace
to be below acceptable limits.
Main PCA.
q
R20/R21 Corrupt
Replace Main PCA.
12
Motor edge lost.
Check blower wire connections. Replace Blower Assembly.
Replace Main PCA.
B
Motor Brake.
Check blower wire connections. Replace Blower Assembly,
Replace Main PCA.
14
Motor Current - The motor current has exceeded its limit.
Check blower wire connections. Replace Blower Assembly.
Replace Main PCA.
15
Rotor Locked - The motor could not be started.
Check Wiring. Replace Blower Assembly. If error persists, replace
Main PCA.
16
Program Sequence - External conditions prevented the proper
Replace Main PCA.
operation of the software.
17
Software Error.
Replace Main PCA. Contact Respironics Product Support.
18
A/D Converter
Replace Main PCA.
κ
2
1011170
REMstar Pro CPAP System
6-15