4.3.2 Alternate and Optional Circuit Accessories
The Respironics alternate accessories listed in Table 4-1 can be used in place of the recommended accessories. The optional accessories can be added to
the patient circuit. Refer to Table 4-1 for changes in the pressure / flow characteristics when using these accessories.
Table 4-1, Patient Circuit Accessory Pressure / Flow
Characteristics
Pressure Drop @ 30 Ipm (cm H,O)
Pressure Drop @ 60 Ipm (cm H,O)
Recommended Patient Circuit
Nasal Mask (disposable, reusable)
0.03
0.09
Profile Lite Nasal Mask (reusable)
0.03
0.09
Simplicity Mask
0.20
'
0.77
Whisper Swivel 1 (reusable)
0.02
0.08
6 ft. (1.83m) Flexible Tubing
0.11
0.30
Alternate Accessories
Spectrum Full Face Mask
0.04
0.10
Plateau Exhalation Valve (reusable)
0.05
0.20
6 ft. Disposable Circuit & Exhalation Port
0.20
0.34
Optional Accessories
REMstar Passover Humidifier
0.00
0.30
REMstar Heated Humidifier
0.00
0.30
King Bacteria Filter (disposable)
0.74
1.65
WARNING!
= Appropriate diagnostic pressures must be determined when alternate or optional accessories are in place. Respironics reusable circuit
accessories are washable but are for single patient use only. They are not intended to be used on multiple patients.
1011170