Pace enhance
From Software release V5.2 the Pace enhance mode can no more be set in the service
mode. Ist accessable in the user mode under "System Setup" ( Allgemeine Einstellungen ),
Pace Enhancement ( Pace Markierung) YES or NO.
3.16
Serial number
The serial number of the device is displayed by pressing the "N" key.
The serial number is only displayed and cannot be changed.
The serial number is entered during the manufacturing process.
After replacing the PCB Control CS, Cİ, during the power up sequence, the serial number
of the device has to be entered.
Terminate with the "X" key.
3.17 ACI-TIPI
Enable or Disable of ACI-TIPI functionality activated by pressing the "O" key, the following
display image appears:
ACI-TIPI
1=
ENABLE
2=
DISABLE
X=
End
NOTE
To DISABLE ACI-TIPI first it must be set to NO in the device user mode.
This must be done in the "Patient Data Setup" screen.
The answer to prompt "Chest Pain" must be set to "NO".
IH
If this is confirmed ACI-TIPI can be enabled in the Service Mode.