Anti-Aliasing Test Pattern; Speaker Test; Keyboard Test - GE Marquette MAC 5000 Field Service Manual

Resting ecg analysis system
Hide thumbs Also See for Marquette MAC 5000:
Table of Contents

Advertisement

Troubleshooting: Diagnostic Tests

Anti-Aliasing Test Pattern

Speaker Test

Keyboard Test

4-18
The second test pattern shows 32 color levels. Check for problems with
the overall pattern. (If the system does not have the color option, various
grey scale patterns display.)
Press the enter key to exit the test.
This test pattern consists of a large square with a series of lines
projecting from the center of the square to the perimeter of the square.
Specifications not currently defined.
Press the enter key to exit the test. Highlight Return and press the enter
key to return to the System Diagnostics menu.
Use the arrow pad to select Loud or Soft. Press the enter key to produce a
loud or soft tone. (The tone level difference is minimal.)
Highlight Return and press the enter key to return to the System
Diagnostics menu.
Press each key and verify that the key is highlighted on the screen and
also displayed at the top of the screen. (It is normal for a dim
background image to remain on the screen when you select the next
key.) The numeric value that displays at the top of the screen is the scan
code representation of the pressed key.
The display shows keys in the upper part of
the screen that are not presently available on
the keyboard.
Check both of the shift keys by pressing each in combination
with a letter to display a capital letter.
Press the center of arrow pad control and verify that the word IN
displays on screen. Press arrows to change the displayed arrow
position. A beep sounds with each arrow press.
Press the shift key and the F6 key to exit the test.
MAC 5000 resting ECG analysis system
2000657-002
NOTE
Revision B