Ametek LAND IQ Series User Manual page 44

Table of Contents

Advertisement

series
CMD
Function
READ
response
IRT
01: IQ_1
Thermo-
meter
LBL
Processor
xx
label
MEM
0000
Memory
#
OBS
0000
Obscu-
ration
IAC
Intl amb
35C
ANC
Analog
1
control
OMN Iout min
500
temp
OMX Iout max
1700
temp
OPT
Iout scale
0000
LSR
Laser
0000
control
OVR
0000
Over-
range
POF
Ppick off
0010
delay
PON
0000
Pick on
delay
A - 6
SET
Range
example
SAIRT
0000-xxxx
0001
SALBL
alpha-
numeric
SAMEM
0000-9999
1101
N/A
0, 1
N/A
-
SAANC
0, 1
1
SAOMN
IRTmin>
700
(max-50)
SAOMX
(IRTmin+50)
>max
1500
SAOPT 1 0, 1
SALSR 1 0, 1
N/A
0, 1
SAPOF
0-100
20
SAPON
0-100
10
Resolution Default
1
0001
1
0000
-
-
-
-
1
1
IRTmin
IRTmax
-
1
0
0
-
-
1
10
1
0
Fixed Spot Thermometer
Comments
Range found using
IF01 command.
NOTE: SET will
reset all processor
configurable
parameters to default
values.
Max 15 characters
Rd resp
0000 RAM=EEPROM
0001 RAM≠EEPROM
Flag is set when user
has changed a config
parameter, but not yet
saved. Processor uses
parameters that will be
lost at power off.
Set arg 1101
Save RAM>EEPROM i.e.
Save configuration 9999
Move EEPROM>RAM
i.e. Restore config
from EEPROM. See
also REFresh for saving
EEPROM>RAM.
0 = not obscured
1 = obscured
Ratio IRTs only. Not
NAMUR 43 compliant.
Internal ambient
(selected IRTs)
thermometer temp.
Analog emissivity
control.
0 = digital
1 = analog
Must also be 50°
below Iout max
Must also be 50°
above Iout min
0 = 0 to 20mA
1 = 4 to 20mA
1 = Laser on
Auto turn off
0 = In range
1 = Over-range
100 steps of 0.1 secs
100 steps of 0.1 secs
User Guide