Fixed Spot Thermometer
Item Description
5
Thermometer
S/N
Address
Type
Label
Units
6
Emissivity
(IQ1, 2 & 5)
Setting
Mode
7
Non greyness
(IQR only)
Setting
Mode
8
Output
Uses an 'AZAR' system (adjustable zero, adjustable range). Allows the user
to set the current output to operate over a reduced span within the operating
range of the instrument.
Maximum
Temperature
Minimum
Temperature
Mode
User Guide
User set?
Function
No
Instrument serial number
Yes
The RS485 address of the thermometer.
To change the address, type the required
address in the text box, then click on
Apply (Item 13). Note that in order to
maintain communication between the
configurator and the thermometer, either
the configurator address must be changed
to match the new thermometer address,
or the 1:1 mode must be selected.
Instrument type
Label or name allocated to the
No
instrument, determined by the user
Yes
°C or °F, selectable by the user
Yes
Yes
Set by the user, range 0.10 to 1.00
Yes
Serial option allows the emissivity to be
set using the digital Setting function.
0-20mA option, reads the analog
emissivity setting directly from the
instrument.
Yes
Set by the user, range 0.800 to 1.250
Yes
Serial option allows the non greyness to
be set using the digital Setting function.
0-20mA option, reads the analog non
greyness setting directly from the
instrument.
Yes
Sets the upper limit of the temperature
span
Yes
Sets the lower limit of the temperature
span
Yes
Sets the output mode to 0-20mA or
4-20mA. Minimum span 50 °C / 90 °F
series
5 - 8