Ultramat 23 Csa General; Ultramat 23 Csa With Option E21; Safety Instructions For The Introduction Of Combustible Gases - Siemens ULTRAMAT 6 Series Compact Operating Instructions

Continuous gas analysis
Table of Contents

Advertisement

2.4.2

ULTRAMAT 23 CSA general

The device is suitable for installation in devices which are permitted by CSA.
2.4.3

ULTRAMAT 23 CSA with option E21

The sample gas exhaust lines must terminate outside the hazardous area at a harmless location.
The sample gas exhaust line can also be returned to the sampling point.
Ensure that the device is operated without pressure at the sample gas outlet.
For the introduction of flammable gases, purging with inert gas or comparable measures to
avoid an explosive atmosphere are recommended, because devices with seals are considered
devices with limited release according to international standards.
Risk of explosion due to accumulation of flammable gases
When the device is installed in an enclosure, an air exchange rate of at least once per hour must
be ensured.
Flammable components in sample gas
If the sample gas could contain flammable components, we recommend that you only use
devices with option - E21.
2.4.4

Safety instructions for the introduction of combustible gases

The maximum permissible sample gas pressure in the analyzer depends on the device version
and is described in the section Technical specifications (Page 51) under the item "Gas inlet
conditions".
You must perform a leak test following all installation and maintenance work affecting the
containment system.
This leak test must be performed at regular intervals.
Series 6 and ULTRAMAT 23
Compact Operating Instructions, 01/2024, A5E45779144002-AB
WARNING
WARNING
Safety instructions
2.4 Analyzers in hazardous areas
19