Siemens ULTRAMAT 6 Series Compact Operating Instructions page 52

Continuous gas analysis
Table of Contents

Advertisement

Technical specifications
7.1 Technical specifications ULTRAMAT 23
Electrical characteristics
Analog outputs
Relay outputs
Binary inputs
Serial interface
AUTOCAL function
Options
Climatic conditions
Permissible ambient temperature
During operation
 
During transportation and storage
Permissible ambient humidity
Permissible ambient pressure
Maximum operating altitude
Pollution degree
Gas inlet conditions
Sample gas pressure
Without pump
 
With pump
Sample gas flow
Sample gas temperature
Sample gas humidity
52
 
1 analog current output per component, 0/2/4/NAMUR ... 20 mA, floating,
max. load 750 Ω
8, with changeover contacts, freely selectable, e.g. for fault, loading ca‐
pacity 24 V AC/DC/1 A, floating, non-sparking
3, dimensioned for 24 V, floating
Pump
AUTOCAL
Synchronization
ELAN (RS485), PROFIBUS-PA/DP as option
Automatic calibration (adjustment) of all channels with ambient air (in case
oxygen is present as measured component) or nitrogen (depending on
measured component),
adjustable cycle time from 1 ... 24 hours. If 0 hours is entered, the cycle is
deactivated.
AUTOCAL cycle must be activated for certified emission measurements
Add-on electronics, with 8 additional digital inputs and 8 additional relay
outputs, for e.g. triggering of automatic calibration, PROFIBUS PA/DP
+5 ... +45 °C (+41 ... +113 °F) (limited when used with H
cf. Assembly/installation (Page 25))
-20 °C ... + 60 °C (-4 ... +140 °F)
< 90% RH (relative humidity) during transportation and storage (Assembly/
installation (Page 25))
600 ... 1200 hPa (absolute)
2000 m above sea level
2 according to IEC 60664-1
before device: On the inlet side, the overpressure of the gas with respect to
ambient pressure must not exceed 45 hPa.
after device: On the outlet side, the gas outlet must be depressurized
against ambient pressure.
before device: Depressurized suction mode
after device: On the outlet side, the gas outlet must be depressurized
against ambient pressure.
72 ... 120 l/h (1.2 .. 2 l/min)
0 ... +50 °C (+32 ... +122 °F)
< 90% RH (relative humidity), non-condensing
Compact Operating Instructions, 01/2024, A5E45779144002-AB
S sensor,
2
Series 6 and ULTRAMAT 23