Technical Specifications Fidamat 6 - Siemens ULTRAMAT 6 Series Compact Operating Instructions

Continuous gas analysis
Table of Contents

Advertisement

Technical specifications
7.8 Technical specifications FIDAMAT 6
Time response
Warm-up time
Delayed display T
(incl. dead time)
90
Damping (electronic time constant)
Dead time (at 1 l/min)
7.8

Technical specifications FIDAMAT 6

Enclosure
Weight
Dimensions (W x D x H)
Electrical characteristics
Power consumption
Fuse values
100 to 120 V
200 to 240 V
Gas inlet conditions
Sample gas pressure
With integrated pump
Without pump
Sample gas temperature
Sample gas humidity
Pressures and flow (FIDAMAT 6 with pump)
Gases
Inlet pressure hPa
(rel.)
 
 
Combustion
2000 ... 4000
gas
Combustion
2000 ... 4000
air
Sample gas
atmospheric
Zero gas
2500 ... 3000
Calibration gas 2500 ... 3000
60
< 30 min, maximum accuracy is achieved after about 2 hours
Approx. 35 s
0 to 100 s, adjustable
Approx. 34 s (because of diffusion to sensor)
Approx. 23 kg
483 mm x 457 mm x 160 mm
Approx. 150 VA during operation
Approx. 350 VA in the warm-up phase
 
4.0 T/250
2.5 T/250
 
600 ... 1100 hPa
< 2000 hPa
0 ... 200 °C
< 90% relative humidity
Operating pressure
hPa (rel.)
Without pump
With pump start-up
start-up
1000 ± 20
420 ± 20
500
-
500 ± 2
-
1500 ± 2
-
1500 ± 2
Flow across FID
ml/min
 
~ 12
~ 320
~ 3
~ 3
~ 3
Compact Operating Instructions, 01/2024, A5E45779144002-AB
Flow through by‐
pass
ml/min
 
-
~ 500
~ 1000
~ 1000
~ 1000
Series 6 and ULTRAMAT 23