Safety instructions
2.6 Measurement of hydrogen in oxygen with Calomat 6E
2.6

Measurement of hydrogen in oxygen with Calomat 6E

NOTICE
Explosive gas mixtures
Gas mixtures of hydrogen H
pressure of 1000 hPa and 20 °C ambient temperature, the lower explosive limit (LEL) lies at 4 %
H
in O
2
the LEL lies at 3.8 % H
In general, a distinction is made between the following cases at the occurrence of explosive
gas mixtures:
• Frequently or constantly explosive - more than 1000 hours/year
• Occasionally explosive - 10 to 1000 hours/year
• Rarely explosive - less than 10 hours/year
Danger of explosion
Measuring explosive gas mixtures with CALOMAT 6E and CALOMAT 6F
• Non-explosive sample gases and gas mixtures below the LEL may be inserted into the
• Rarely or occasionally explosive gas mixtures must only be introduced into the gas
• Frequently or continuously explosive gas mixtures must not be introduced into the gas
• Gas mixtures above the upper explosive limit: Assess the introduction individually in
NOTICE
Checking leak tightness of the gas path
In case of leakages of the gas path (containment system, CS), hydrogen (H
can concentrate in the gas analyzer to form an explosive mixture. Only after checking for a
suitable leak tightness of the gas path, may these gases be fed into the gas analyzer.
CALOMAT 6E rack unit
The enclosure of the gas analyzer has to be equipped with sufficient openings so that a gas
exchange with the environment takes place and an explosive mixture cannot build up in the
device.
22
, the upper explosive limit (UEL) at 95.2 % H
2
in O
2
DANGER
CALOMAT 6. Flame arrestors can be omitted. However, it must be guaranteed that the
sample gas in the gas path (containment system, CS) always remains below the lower
explosive limit!
analyzer if the inputs and outputs for sample gas have each been provided with a flame
arrestor. The flame arrestors must meet the safety requirements for the combustible
medium.
analyzer.
accordance with a separate certificate based on on-site acceptance testing.
in oxygen O
are explosive in a wide range. At an atmospheric
2
2
2
, the UEL lies at 95.4 % H
2
Compact Operating Instructions, 01/2024, A5E45779144002-AB
in O
. At an ambient temperature of 80 °C
2
in O
.
2
2
) and oxygen (O
2
Series 6 and ULTRAMAT 23
)
2